91大神

91大神Water Technologies & Solutions

Sievers UPW Boron Analyzer

Continuous, online monitoring of boron with Sievers UPW Boron Analyzer

The Sievers Boron Analyzers enable continuous, unattended online monitoring of ultra-low levels of boron in deionized ultrapure water applications. It's able to run 10 samples per hour and has a limit of detection (LOD) of 15 ppt boron. The Sievers Boron Analyzer is designed to meet the needs of facility managers, DI water engineers, and fab process engineers.

Easy to operate and maintain

The Sievers Boron Analyzer predicts mixed-bed exhaustion, optimizes EDI performance and controls polishing loop boron levels. The boron analyzer detects an increase in boron prior to silica release, making it useful to prevent the leakage of silica into your ultrapure water. The Sievers Boron Analyzer can significantly reduce operating expenses and maintain quality in UPW systems.

Features and Benefits

Boron Features and Benefits

Enhanced reliability and greater uptime in challenging applications ?

  • Continuous, on-line monitoring as low as 15 ppt B ?
  • Multi-stream options – up to four sample streams ?
  • Extensive suite of outputs – four isolated 4–20 mA outputs, one isolated RS-232, one universal parallel printer port, two alarm outputs, and four isolated binary outputs?
  • Minimal maintenance and consumables ?
  • Accessories kit includes accessories needed for setup?

Applications?

  • Semiconductor Ultrapure Process Water?
  • Ultrapure Water Power Applications

Specifications

Boron Specifications

Operating Specifications

Concentration Range 0.015 to 20 ppb B
Limit of Detection (LOD) 0.015 ppb B
Accuracy ± 0.01 ppb B <0.05 ppb B
± 15% >0.05 ppb B
Precision ± 3% or 0.005 ppb B, whichever is greater
Minimum Sample
Resistivity

15 megaohms–cm

Sample Temperature 15°-40°颁
Sample Flow Rate 0.1 –0.5 L/min
Ambient Temperature 10°颁–40°颁
Sample Pressure 20–100 psig (138-689 kPa)
Calibration Stability Typically?12 months
Sample Particulates 15 micron sample pre-filtration
Waste Gravity Drain (no back pressure)

Instrument Specifications

Inputs 5 isolated binary inputs
Outputs 4 isolated 4-20mA outputs, 1 RS-232 (isolated), 1 universal parallel printer port, 2 alarm outputs, 4 isolated binary outputs
Dimensions

20.625 H x?17.25 W x 12.25 D inches (53 x 44 x 31 cm)

Weight 48 lbs (22kg)
? ?

Consumables

Boron Reagent 90 days @ 12 minute sampling rate
45 days @ 6.5 minute sampling rate
Reagent Column 12 months
Zero Column 12 months